XENAMATIC-2000N and 3000N -
The Xenon 133(127) Gas Delivery System
|
|
XENAMATIC-2000N
and 3000N series of xenon rebreathing devices are designed to provide both
the user and patient with maximum flexibility and safety when performing
lung ventilation studies with either Xe133 or Xe127.
Resistance to normal breathing is the lowest in the industry. This, coupled
with an O2 monitor and automatic O2 replenishment,
means that it is well tolerated by most patients. The XENAMATIC-3000N has a
universal interface which enables it to be used with all ventilators, thus
allowing studies on patients that require breathing assistance. The
activated charcoal xenon Trap is very large, and air flow is controlled to
provide exceptionally long life.
EASE OF USE,
LOW BREATHING RESISTANCE, AUTOMATIC O2 REPLENISHMENT, LONG TRAP
LIFE, AND VENTILATOR PATIENT CAPABILITY MEANS A BETTER SYSTEM.
|
|
Price
|
|
Specifications
|
Electronic
Control - The XENAMATIC-3000N
series is equipped with a state-of-the-art combination digital display and
membrane switch control panel. All valves in the breathing circuit are
electrically operated to assure rapid response and have a 0.75 inch orifice
for low breathing resistance.
|
FUNCTION MODES -The following functions are provided from the
membrane switch control panel.
|
Room Air - Patient inhales from and exhales into the room
through the XENAMATIC-3000N plumbing system.
|
Rebreathe - Patient breathes in the closed loop, coming to
equilibrium with the xenon stored in the system. REBREATHE will start
automatically at end tidal volume so that the first breath is always an
inhalation. While in REBREATHE, CO2 is removed from the closed
loop, and as O2 is consumed, it is automatically replaced.
|
Washout - Patient inhales room air through the
XENAMATIC-3000N and exhales into the Expandable Interface. The Xenon Trap
pump is activated and draws the air- xenon mixture from the Expandable
Interface into the charcoal Trap. The control of the rate of flow and the
volume assures Xenon trapping and greatly improves trap life.
|
End Study - Patient is removed and the Xenon Trap pump
continues to run until a vacuum sensor indicates the Expandable Interface
is empty. The Trap pump then shuts off and returns the unit to the ROOM AIR
mode.
|
|
Note: The
above modes may be executed through in sequence with our optional REMOTE
CONTROL..
|
|
Auto Fill - Automatically fills the internal 5 liter
breathing bag to approximately 3.5 liters, so that there is always an
adequate volume for the first inhalation, even for a single
breath-equilibrium-washout technique.
|
Flush System - Flushes room air through the system into the
Expandable Interface and charcoal Trap to clean all radioactivity out of
the system.
|
|
Other
Functions
|
|
Alarm Audio - ON/OFF switch for the audio portion of the alarm
indicator system. Visual alarm indicator is always active.
|
|
The
following conditions will activate the alarm system and will be indicated
in the status window:
|
|
Fill Err - Filling errors during AUTO FILL or O2 problems
during REBREATHE.
|
Rstrictn - Airway pressure or vacuum exceeds 5cm H2O.
|
Xe Trap - Trap exhaust contains more than 0.01 mCi/ml of
radioactivity.
|
|
Other error
messages are provided. See manual for complete information.
|
|
Activity Used
- Displays the activity used
during a patient study by subtracting the [end volume X concentration] from
the [start volume X concentration]. This represents the "patient
dose".
|
Ventilator
Control- The Model 3000N may be
connected to any ventilator to study patients who require breathing
assistance. Xenon never comes into contact with the ventilator.
|
O2
Monitor & Auto O2 Control- The XENAMATIC-3000N series contains a paramagnetic
O2 sensor which reads out in % O2 and allows a user
selectable maintenance setting throughout REBREATHE. No periodic user
service is required.
|
Xenon
Activity- The XENAMATIC-3000N
provides a readout of the concentration of Xenon in either mCi/L or GBq/L.
|
Bacteriological
Filter- The XENAMATIC-3000N uses
an external single use disposable bacteriological filter.
|
CO2
Absorber- Large capacity,
in-line, soda lime CO2 absorber is provided.
|
Trapping
System- The overall Xenon Trap
system is the most efficient in the industry. Xenon Trapping is most
effective at flow rates below 10L/min. Patients breathe at 15L/min. Our
Trap pump has a flow rate of 7-8L/min. The flow rate difference is stored
in a 150L Expandable Interface. At END STUDY, the Trap pump continues to
operate until the Expandable Interface is empty and then automatically
shuts off. Air flow over charcoal is the single biggest cause of Trap
failure. This is automatically kept to a minimum in the XENAMATIC.
|
|
DIMENSIONS:
|
30"w x 24.5"d x 53.5"h (78cm x 62cm x 136cm)
|
WEIGHT:
|
500 lbs (225 kg)
|
ELECTRICAL:
|
115V, 60Hz, 1.5Amp or 230V, 50Hz, 1.3Amp
|
SAFETY
APPROVAL:
|
MET-NRTL, MET-C; (UL 2601-1, CSA 601.1)
|
|
|
XENAMATIC
2000N
|
Please call for current price Note: Not for ventilator patients
|
XENAMATIC
3000N
|
Please call for current price
|
|
O2 Monitor
and Auto O2 replenishment, ventilator interface, retractable
arm, total lead shielding 1/8" (1/4" Trap area) with optional
shielding for Xe127, large capacity CO2 Trap, 20 ft -
3" diameter charcoal cartridge pack with Expandable Interface,
moisture absorber, and built-in exhaust port monitor/alarm.
|
|
Accessories and
Options(Please contact us for current prices)
|
|
Breathing
Port Adapter*
|
|
CO2
Absorber (2 refills) *
|
|
Bac-T
Filter, Disposable*
|
|
Desiccant
(Silica gel) 2lbs *
|
|
Face Mask, Adult
*
|
|
Face Mask
Harness, Adult*
|
|
Face Mask
Tubing (8")*
|
|
Xenon Gun
Adapter*
|
|
Xenon Trap
Charcoal Cartridge *
|
|
Face Mask, Infant
|
|
Face Mask Tubing (5")
|
|
Remote Controller
|
|
|
|
Lead
Shielding for Xe127
|
|
CBF
Modification (Model 4000)
|
|
230V, 50Hz
Conversion
|
|
|
*One of each (Five
Bac-T filters included)is
included with the purchase of the XENAMATIC-3000N
|
|
WARRANTY:
One Year Parts and Labor
|
TERMS: Net
30; F.O.B. Houston
|
|